Emerson PN 51-FCL-1056 instruction manual ZeroBufferCalCal, Auto

Page 53

MODEL FCL-1056

SECTION 6.0

 

CALIBRATION

5.USER ENTERED SLOPE AND OFFSET. If the slope and offset are known from other measurements, they can be directly entered in the analyzer. Enter the slope as a positive number corrected to 25ºC. To calculate the slope at 25ºC from the slope at temperature tºC, use the equation:

slope at 25ºC = (slope at tºC)

298

tºC + 273

To calculate the offset use the following equation. The offset can be either positive or negative.

offset = mVbuffer - (pHbuffer - 7.00)(slope at 25ºC)

6.STABILITY SETTING. During automatic calibration, the analyzer measures noise and drift and does not accept calibration data until readings are stable. Calibration data will be accepted as soon as the pH reading is constant to within the factory-set limits of less than 0.02 pH change in 10 seconds. The stability settings are programmable.

6.4.3 Procedure-Auto Calibration

1.Obtain two buffer solutions. Ideally, the buffer pH values should bracket the range of pH to be measured.

2.Remove the sensor from the flow cell. If the process and buffer temperatures are appreciably different, place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has reached the buffer temperature.

Calibrate?

Sensor 1

SensorSensor22

Output 1

Output 2

3.Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER.

4.Choose the sensor you wish to calibrate. Sensor 1 is the chlorine sensor. Sensor 2 is the pH sensor.

S2 Calibration

pHpH

ORP

Redox Temperature

S2 pH Cal

ZeroBufferCalCal

Standardize

Slope: 56.19 mV/pH

Offset: 2mV

5. Choose pH.

6.Choose Buffer Cal

S2 pH Buffer Cal

Auto

Manual

7. Choose Auto.

43

Image 53
Contents FCL with 1056 Analyzer Essential Instructions Free Chlorine EnglishPpm mg/L ORPS1 Manual pH Temp Units S1 Free Cl PH Correction ManualLive/Continous Sensor 2 pH Menu TreeCalibrate Sensor 1 Free chlorine Temperature AlarmsDiagnostic Setup Reset AnalyzerRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DApplications Section Description and SpecificationsFeatures Specifications Analyzer Specifications GeneralSpecifications Sensor Ordering Information AccessoriesComponent Parts PH Correction required selectionThis page left blank intentionally Unpacking and Inspection Section InstallationFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstalling the Sensors InstallationGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally POWER, ALARM, and Output Wiring Section WiringPower Analog output wiringAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Calibrate Programming the ANALYZER-TUTORIALOutputs RangeO1 S1 20mA 08.50 ppm Output Range O1 S1 4mA 0.000 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityHold Using HoldMain Format Configuring the Main DisplayLanguage English Contrast Format, Language, Warning, and ContrastGeneral Default SettingsSection Programming the Analyzer Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsAssign ConfigureProgramOutput Procedure Ranging Outputs Configure Simulate Output ConfigureOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfiguring Alarms and Assigning Setpoints Configure SimulateRangeSimulate OutputSection Programming the Analyzer Alarm AlarmsConfigure/Setpoint Alarm Simulate Alarm Alarms Configure/SetpointSimulate Don’t SimulateProcedure Synchronizing Timers Alarms Configure/Setpoint SimulateSynch Timers Yes Definitions pH/ORP Configuring the MeasurementDefinitions Chlorine Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsTemperature Program Outputs Alarms MeasurementConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompSetting UP Diagnostics Program Alarms Measurement TemperatureSecurity SecurityDiagnostic Setup Program Measurement Temperature SecurityGI Fault High 1500MΩ Procedure Setting Up DiagnosticsProgram Temperature Security Diagnostics Resetting the AnalyzerReset Analyzer Reset AnalyzerIntroduction Calibrating TemperatureSection Calibration Calibrate SensorSensor11 Output ProcedureS1 Calibration S1 Calibration + 25.0CCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalSensor Sensor OutputCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCalStable Time 07.01 pH10.01 pH Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer StandardizeStandardize Redox Temperature S2 pH Cal Buffer CalSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset MA Output Cal Meter 000 mA Trim Complete Calibration Analog OutputsOutput This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceReplacing the electrolyte solution and membrane Cleaning the membraneChlorine Sensor GeneralPH Sensor Cleaning the SensorOther Maintenance Constant Head Flow Controller Cleaning the flow controllerWeight Weight Using the Diagnostic Feature Section TroubleshootingFaults OverviewHardware Error Troubleshooting When a Fault Message is ShowingFault message Explanation Section Sensor ADC Error Sensor CPU ErrorSensor Incompatible Sensor RTD OpenGlass Z Too High Sensor RTD Out of RangeReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Readings drift Sensor can be calibrated, but the current is too lowProcess readings are erratic Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Simulating Inputs Chlorine Troubleshooting When no Error Message is Showing GeneralAnod CathSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage ASIA-PACIFIC Americas HeadquartersEurope GermanyWarranty Specifications subject to change without notice