Emerson PN 51-FCL-1056 instruction manual Calibrate?, Sensor Output, PFree Chlorine

Page 51

MODEL FCL-1056

SECTION 6.0

 

CALIBRATION

6.3.3 Procedure-Calibrating the Sensor

1.Place the chlorine sensor in the chlorine flow cell. If continuous (live) pH correction is being used, calibrate the pH sensor (section 6.4) and place it in the pH flow cell. If manual pH correction is being used, measure the pH of the sample and enter the value. See section 5.5. Adjust the sample flow until water overflows the cen- ter tube of the constant head flow controller.

2.Adjust the chlorine concentration until it is near the upper end of the operating range. Wait until the analzyer reading is stable before starting the calibration.

Calibrate?

Sensor 1

Sensor 2

Output 1

Output 2

3.Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER.

4.Choose the sensor you wish to calibrate. Sensor 1 is the chlorine sensor. Sensor 2 (if present) is the pH sensor.

S1 Calibration

pFree Chlorine

Temperature

S1 Calibration ZeroZeroCalCal

In Process Cal

5.Choose Free Chlorine.

6.Choose In Process Cal

S1 Enter Value

1 0.00 ppm

S1 InProcess Cal Calibration Error

Press Exit

7.Follow the screen prompts: Once the reading is stable, press ENTER. Take the sample and press ENTER. At this point, the analyzer will store the present sensor current and temperature and use those values in the calibration.

Determine the free chlorine concentration in the sample and enter the value in the screen shown at left. See Section 6.3.1 for sampling and testing precautions.

8.If the calibration is successful, the live reading will change to the value entered in step 7 and the display will return to the screen in step 6.

If the sensitivity is too far outside the range of expected values, the calibration error screen shown at left will appear. The analyzer will not update the calibration. For troubleshooting assistance, see Section 9.5.

9. To return to the main display, press MENU then EXIT.

41

Image 51
Contents FCL with 1056 Analyzer Essential Instructions ORP EnglishFree Chlorine Ppm mg/LS1 Free Cl PH Correction Manual Live/ContinousS1 Manual pH Temp Units Menu Tree Calibrate Sensor 1 Free chlorineSensor 2 pH Reset Analyzer AlarmsTemperature Diagnostic SetupRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DSection Description and Specifications FeaturesApplications Specifications General Specifications SensorSpecifications Analyzer PH Correction required selection AccessoriesOrdering Information Component PartsThis page left blank intentionally FCL-02 free chlorine with continuous pH correction Section InstallationUnpacking and Inspection FCL-01 free chlorine without continuous pH correctionMounting, Inlet, and Drain Connections InstallationInstalling the Sensors General InformationModel FCL-01 This page left blank intentionally Analog output wiring Section WiringPOWER, ALARM, and Output Wiring PowerAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Range Programming the ANALYZER-TUTORIALCalibrate OutputsSecurity Output Range O1 S1 4mA 0.000 ppmO1 S1 20mA 08.50 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CHold Using HoldFormat, Language, Warning, and Contrast Configuring the Main DisplayMain Format Language English ContrastDefault Settings Section Programming the AnalyzerGeneral Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsConfigure ProgramOutputAssign O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0C Configure Simulate Output ConfigureProcedure Ranging Outputs Output Output RangeOutput Configure SimulateConfiguring Alarms and Assigning Setpoints RangeSimulateSection Programming the Analyzer Alarms Configure/SetpointAlarm Don’t Simulate Alarms Configure/SetpointAlarm Simulate Alarm SimulateAlarms Configure/Setpoint Simulate Synch Timers YesProcedure Synchronizing Timers Configuring the Measurement Definitions ChlorineDefinitions pH/ORP Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsS1 Temp Comp Auto S2 Temp Comp Program Outputs Alarms MeasurementTemperature Configuring Security SettingsSecurity Program Alarms Measurement TemperatureSetting UP Diagnostics SecurityProcedure Setting Up Diagnostics Program Measurement Temperature SecurityDiagnostic Setup GI Fault High 1500MΩReset Analyzer Resetting the AnalyzerProgram Temperature Security Diagnostics Reset AnalyzerCalibrating Temperature Section CalibrationIntroduction S1 Calibration + 25.0C ProcedureCalibrate SensorSensor11 Output S1 CalibrationCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationSensor Output Temperature S1 Calibration ZeroZeroCalCal Process CalCalibrate? SensorCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCal07.01 pH 10.01 pHStable Time Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer S1 Enter Value 00 pH Redox Temperature S2 pH Cal Buffer CalStandardizeStandardize Slope 56.19 mV/pH OffsetProcedure-Entering a Known Slope and Offset Calibration Analog Outputs OutputMA Output Cal Meter 000 mA Trim Complete This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceGeneral Cleaning the membraneReplacing the electrolyte solution and membrane Chlorine SensorCleaning the Sensor Other MaintenancePH Sensor Constant Head Flow Controller Cleaning the flow controllerWeight Weight Overview Section TroubleshootingUsing the Diagnostic Feature FaultsTroubleshooting When a Fault Message is Showing Fault message Explanation SectionHardware Error Sensor RTD Open Sensor CPU ErrorSensor ADC Error Sensor IncompatibleBroken Glass Sensor RTD Out of RangeGlass Z Too High Reference Impedance Too HighTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Sensor can be calibrated, but the current is too low Process readings are erraticReadings drift Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Cath Troubleshooting When no Error Message is Showing GeneralSimulating Inputs Chlorine AnodSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage Germany Americas HeadquartersASIA-PACIFIC EuropeWarranty Specifications subject to change without notice