MODEL | SECTION 9.0 |
| TROUBLESHOOTING |
9.5.6 Sensor does not respond to changes in chlorine level.
1.Is the grab sample test accurate? Is the grab sample representative of the sample flowing to the sensor?
2.Is sample flowing past the sensor? Be sure the liquid level in the constant head sampler is level with the central overflow tube and that excess sample is flowing down the tube. If necessary, disassemble and clean the over flow sampler. See Section 8.4.
3.Is the pH compensation correct? If the analyzer is using manual pH correction, verify that the pH value in the analyzer equals the actual pH to within ±0.1 pH. If the analyzer is using automatic pH correction, check the calibration of the pH sensor.
4.Is the membrane clean? Clean the membrane and replace it if necessary. Check that the holes at the base of the cathode stem are open. Use a straightened paper clip to clear blockages. See step 4 in Section 9.5.2. Replace the electrolyte solution.
5.Replace the sensor.
9.5.7 Chlorine readings spike following sudden changes in pH (automatic pH correction).
Changes in pH alter the relative amounts of hypochlorous acid (HOCl) and hypochlorite ion
9.5.8 Chlorine readings are too low.
1.Was the sample tested as soon as it was taken? Chlorine solutions are unstable. Test the sample immediately after collecting it. Avoid exposing the sample to sunlight.
2.Low readings can be caused by zeroing the sensor before the residual current has reached a stable minimum value. Residual current is the current the sensor generates even when no chlorine is in the sample. It is also called the zero current. Because the residual current is subtracted from subsequent measured currents, zeroing before the current is a minimum can lead to low results.
Example: The true residual current for a free chlorine sensor is 4 nA, and the sensitivity is 350 nA/ppm. Assume the measured current is 200 nA. The true concentration is
3.Sensor response depends on flow. If the flow is too low, readings will be low and flow sensitive. Verify that the chlorine sensor is installed in the correct flow cell. See Figures
66