Emerson PN 51-FCL-1056 instruction manual Simulating Inputs pH, Simulating pH input

Page 81

MODEL FCL-1056

9.9 SIMULATING INPUTS — pH

9.9.1 General

This section describes how to simulate a pH input into the analyzer. To simulate a pH measurement, connect a standard millivolt source to the analyzer. If the analyzer is working properly, it will accurately measure the input voltage and convert it to pH.

9.9.2 Simulating pH input.

1.Set automatic temperature correction to manual and set manual temperature to 25°C. Turn off solu- tion temperature correction. See Section 5.6

2.Disconnect the sensor and connect a jumper wire between the IN REFERENCE and IN pH terminals.

3.Press DIAG and choose sensor 2 (pH). The input voltage should be 0 mV and the pH should be 7.00. Because calibration data stored in the analyzer may be offsetting the input voltage, the displayed pH may not be exactly 7.00.

4.If a standard millivolt source is available, discon- nect the jumper wire between IN REFERENCE and IN pH and connect the voltage source as shown in Figure 9-4. Be sure to jumper the IN REFERENCE and GND SOL terminals.

5.Calibrate the analyzer using the procedure in Section 6.4.4. Use 0.0 mV for Buffer 1 (pH 7.00) and -177.4 mV for Buffer 2 (pH 10.00). If the ana- lyzer is working properly it should accept the cali- bration. The slope should be 59.16 mV/pH and the offset should be zero.

6.To check linearity, return to the main display and the pH/temperature/mV screen. Set the voltage source to the values shown in the table and verify that the pH and millivolt readings match the values in the table.

 

Voltage (mV)

pH (at 25°)

 

 

295.8

2.00

 

 

177.5

4.00

 

 

59.2

6.00

 

 

-59.2

8.00

 

 

-177.5

10.00

 

-295.8

12.00

 

SECTION 9.0

TROUBLESHOOTING

FIGURE 9-4. Simulating pH Input.

71

Image 81
Contents FCL with 1056 Analyzer Essential Instructions Free Chlorine EnglishPpm mg/L ORPS1 Free Cl PH Correction Manual Live/ContinousS1 Manual pH Temp Units Menu Tree Calibrate Sensor 1 Free chlorineSensor 2 pH Temperature AlarmsDiagnostic Setup Reset AnalyzerRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DSection Description and Specifications FeaturesApplications Specifications General Specifications SensorSpecifications Analyzer Ordering Information AccessoriesComponent Parts PH Correction required selectionThis page left blank intentionally Unpacking and Inspection Section InstallationFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstalling the Sensors InstallationGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally POWER, ALARM, and Output Wiring Section WiringPower Analog output wiringAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Calibrate Programming the ANALYZER-TUTORIALOutputs RangeO1 S1 20mA 08.50 ppm Output Range O1 S1 4mA 0.000 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityHold Using HoldMain Format Configuring the Main DisplayLanguage English Contrast Format, Language, Warning, and ContrastDefault Settings Section Programming the AnalyzerGeneral Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsConfigure ProgramOutputAssign Procedure Ranging Outputs Configure Simulate Output ConfigureOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfiguring Alarms and Assigning Setpoints Configure SimulateRangeSimulate OutputSection Programming the Analyzer Alarms Configure/SetpointAlarm Alarm Simulate Alarm Alarms Configure/SetpointSimulate Don’t SimulateAlarms Configure/Setpoint Simulate Synch Timers YesProcedure Synchronizing Timers Configuring the Measurement Definitions ChlorineDefinitions pH/ORP Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsTemperature Program Outputs Alarms MeasurementConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompSetting UP Diagnostics Program Alarms Measurement TemperatureSecurity SecurityDiagnostic Setup Program Measurement Temperature SecurityGI Fault High 1500MΩ Procedure Setting Up DiagnosticsProgram Temperature Security Diagnostics Resetting the AnalyzerReset Analyzer Reset AnalyzerCalibrating Temperature Section CalibrationIntroduction Calibrate SensorSensor11 Output ProcedureS1 Calibration S1 Calibration + 25.0CCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalSensor Sensor OutputCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCal07.01 pH 10.01 pHStable Time Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer StandardizeStandardize Redox Temperature S2 pH Cal Buffer CalSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset Calibration Analog Outputs OutputMA Output Cal Meter 000 mA Trim Complete This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceReplacing the electrolyte solution and membrane Cleaning the membraneChlorine Sensor GeneralCleaning the Sensor Other MaintenancePH Sensor Constant Head Flow Controller Cleaning the flow controllerWeight Weight Using the Diagnostic Feature Section TroubleshootingFaults OverviewTroubleshooting When a Fault Message is Showing Fault message Explanation SectionHardware Error Sensor ADC Error Sensor CPU ErrorSensor Incompatible Sensor RTD OpenGlass Z Too High Sensor RTD Out of RangeReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Sensor can be calibrated, but the current is too low Process readings are erraticReadings drift Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Simulating Inputs Chlorine Troubleshooting When no Error Message is Showing GeneralAnod CathSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage ASIA-PACIFIC Americas HeadquartersEurope GermanyWarranty Specifications subject to change without notice