Emerson 1056 Sensor Incompatible, Sensor CPU Error, Sensor RTD Open, Sensor ADC Error

Page 71

MODEL FCL-1056

SECTION 9.0

 

TROUBLESHOOTING

9.3.4 Sensor Incompatible

This error message means that the sensor board software is not supported by the main board software. Either the sensor board or the main board software is too old.

Replace the main board with one compatible with the sensor board. Call the factory for assistance. You will be asked for the main and sensor board software revision numbers. To read the main board software revision, press the DIAG key and scroll down until Instr SW Ver is showing. To view the sensor board software revision, press the DIAG key, choose the appropriate sensor, and scroll down until Board SW Ver is showing. The main board can be replaced only at the factory.

9.3.5 Sensor CPU Error

This message means the sensor board software is corrupted.

1.Cycle the power off then on.

2.If cycling the power does not help, call the factory. The sensor board must be replaced.

9.3.6 Sensor RTD Open

The chlorine and pH sensors used in the FCL contain a Pt 100 RTD (resistance temperature device) for measuring temperature. Sensor RTD open means the temperature measuring circuit is open.

1.Confirm that the sensor RTD wires are properly connected.

2.Confirm that the Varipol connector is properly seated.

3.Disconnect the sensor from the cable and use an ohmmeter to check the resistance across the RTD. See Figures 9-1 and 9-2. At room temperature it should be about 110Ω. If the resistance is very high, the RTD has failed and the sensor must be replaced. If the resistance is okay, connect the Variopol cable to the sensor and disconnect the three RTD wires at the analyzer. Measure the resistance across the red and white RTD leads. If the resistance is very high, the problem is with the VP cable, and it must be replaced.

FIGURE 9-1. Pin Out Diagram for Model 499ACL-01-VP

Sensor (top view of connector end of sensor)

9.3.7 Sensor 1 Not Detected

FIGURE 9-2. Pin Out Diagram for Model 3900VP Sensor (top view of connector end of sensor)

The ribbon cable from sensor 1 (chlorine) board must be plugged into the sensor 1 plug. See Figure 3-1 for the location of the sensor board connectors.

1.Confirm that the ribbon cable connecting sensor 1 (chlorine) board to the main board is plugged into the Sensor 1 connector on the main board.

2.Confirm that the ribbon cable is seated at both ends.

9.3.8 Sensor Factory Data, Sensor Board User Data, and Sensor EEPROM Write Error

These messages mean factory eeprom data or user eeprom data on the sensor board is corrupted or the CPU on the sensor board is bad.

1.Cycle power off then on.

2.Replace the sensor board.

9.3.9 Sensor ADC Error

There is a bad component on the sensor board. The sensor board must be replaced.

61

Image 71
Contents FCL with 1056 Analyzer Essential Instructions ORP EnglishFree Chlorine Ppm mg/LS1 Manual pH Temp Units S1 Free Cl PH Correction ManualLive/Continous Sensor 2 pH Menu TreeCalibrate Sensor 1 Free chlorine Reset Analyzer AlarmsTemperature Diagnostic SetupRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DApplications Section Description and SpecificationsFeatures Specifications Analyzer Specifications GeneralSpecifications Sensor PH Correction required selection AccessoriesOrdering Information Component PartsThis page left blank intentionally FCL-02 free chlorine with continuous pH correction Section InstallationUnpacking and Inspection FCL-01 free chlorine without continuous pH correctionMounting, Inlet, and Drain Connections InstallationInstalling the Sensors General InformationModel FCL-01 This page left blank intentionally Analog output wiring Section WiringPOWER, ALARM, and Output Wiring PowerAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Range Programming the ANALYZER-TUTORIALCalibrate OutputsSecurity Output Range O1 S1 4mA 0.000 ppmO1 S1 20mA 08.50 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CHold Using HoldFormat, Language, Warning, and Contrast Configuring the Main DisplayMain Format Language English ContrastGeneral Default SettingsSection Programming the Analyzer Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsAssign ConfigureProgramOutput O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0C Configure Simulate Output ConfigureProcedure Ranging Outputs Output Output RangeOutput Configure SimulateConfiguring Alarms and Assigning Setpoints RangeSimulateSection Programming the Analyzer Alarm AlarmsConfigure/Setpoint Don’t Simulate Alarms Configure/SetpointAlarm Simulate Alarm SimulateProcedure Synchronizing Timers Alarms Configure/Setpoint SimulateSynch Timers Yes Definitions pH/ORP Configuring the MeasurementDefinitions Chlorine Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsS1 Temp Comp Auto S2 Temp Comp Program Outputs Alarms MeasurementTemperature Configuring Security SettingsSecurity Program Alarms Measurement TemperatureSetting UP Diagnostics SecurityProcedure Setting Up Diagnostics Program Measurement Temperature SecurityDiagnostic Setup GI Fault High 1500MΩReset Analyzer Resetting the AnalyzerProgram Temperature Security Diagnostics Reset AnalyzerIntroduction Calibrating TemperatureSection Calibration S1 Calibration + 25.0C ProcedureCalibrate SensorSensor11 Output S1 CalibrationCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationSensor Output Temperature S1 Calibration ZeroZeroCalCal Process CalCalibrate? SensorCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCalStable Time 07.01 pH10.01 pH Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer S1 Enter Value 00 pH Redox Temperature S2 pH Cal Buffer CalStandardizeStandardize Slope 56.19 mV/pH OffsetProcedure-Entering a Known Slope and Offset MA Output Cal Meter 000 mA Trim Complete Calibration Analog OutputsOutput This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceGeneral Cleaning the membraneReplacing the electrolyte solution and membrane Chlorine SensorPH Sensor Cleaning the SensorOther Maintenance Constant Head Flow Controller Cleaning the flow controllerWeight Weight Overview Section TroubleshootingUsing the Diagnostic Feature FaultsHardware Error Troubleshooting When a Fault Message is ShowingFault message Explanation Section Sensor RTD Open Sensor CPU ErrorSensor ADC Error Sensor IncompatibleBroken Glass Sensor RTD Out of RangeGlass Z Too High Reference Impedance Too HighTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Readings drift Sensor can be calibrated, but the current is too lowProcess readings are erratic Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Cath Troubleshooting When no Error Message is Showing GeneralSimulating Inputs Chlorine AnodSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage Germany Americas HeadquartersASIA-PACIFIC EuropeWarranty Specifications subject to change without notice