Error Messages

 

 

Execution errors

 

 

 

Error

Error Description

description/explanation/examples

Number

 

 

–222

Data out of range

Indicates that a legal program data element was

 

parsed but could not be executed because the inter-

 

 

 

 

preted value was outside the legal range as defined

 

 

by the counter (see IEEE-488.2, 11.5.1.1.5.).

 

Data out of range;

The expression was too large for the internal float-

 

exponent too large

ing-point format.

 

Data out of range;

Data below minimum for this function/parameter.

 

below minimum

 

 

Data out of range;

Data above maximum for this function/ parameter.

 

above maximum

 

 

Data out of range;

A number outside 0 to 19 was given for the save/re-

 

(Save/recall mem-

call memory.

 

ory number)

 

–223

Too much data

Indicates that a legal program data element of block,

Too much data;

expression, or string type received that contained

 

 

PUD string too

more data than the counter could handle due to

 

long

memory or related counter-specific requirements.

 

Too much

 

 

data;String or block

 

 

too long

 

–224

Illegal parameter

Used where exact value, from a list of possible val-

value

ues, was expected.

 

–230

Data corrupt or

Possibly invalid data; new reading started but not

stale

completed since last access.

 

–231

Data questionable

 

Data questionable;

One or more data elements sent with a MEASure or

 

 

one or more data el-

CONFigure command was ignored by the counter.

 

ements ignored

 

–240

Hardware error

Indicates that a legal program command or query

 

could not be executed because of a hardware prob-

 

 

 

 

lem in the counter. Definition of what constitutes a

 

 

hardware problem is completely device specific. This

 

 

error message is used when the counter cannot de-

 

 

tect the more specific errors described for errors

 

 

–241 through –249.

8-8 Error Code -222 to -240

Page 122
Image 122