Error Messages

 

 

 

 

 

 

 

Command Errors

 

 

 

 

 

Error

Error Description

Description/Explanation/Examples

 

Number

 

 

 

–105

GET not allowed

A Group Execute Trigger was received within a pro-

 

 

gram message (see IEEE-488.2, 7.7).

 

 

 

 

–108

Parameter not al-

More parameters were received than expected for

 

lowed

the header; for example, the EMC common com-

 

 

 

 

 

mand accepts only one parameter, so receiving

 

 

 

EMC0,,1 is not allowed.

 

–109

Missing parameter

Fewer parameters were received than required for

 

 

the header; for example, the EMC common com-

 

 

 

 

 

 

mand requires one parameter, so receiving EMC is

 

 

 

not allowed.

 

–110

Command header

An error was detected in the header. This error

 

error

message is used when the counter cannot detect

 

 

 

 

 

the more specific errors described for errors –111

 

 

 

though –119.

 

–111

Header separator

A character that is not a legal header separator was

 

error

encountered while parsing the header; for example,

 

 

 

 

 

no space followed the header, thus GMC"MACRO"

 

 

 

is an error.

 

–112

Program mnemonic

The header contains more than 12 characters (see

 

too long

IEEE-488.2, 7.6.1.4.1).

 

 

 

–113

Undefined header

The header is syntactically correct, but it is unde-

 

 

fined for this specific counter; for example, XYZ is

 

 

 

 

 

 

not defined for any device.

 

–114

Header suffix out of

Indicates that a non-header character has been en-

 

range

countered in what the parser expects is a header el-

 

 

 

 

 

ement.

 

–120

Numeric data error

This error, as well as errors –121 through –129, are

 

Numeric data error;

generated when parsing a data element that ap-

 

 

 

 

overflow from con-

pears to be of a numeric type. This particular error

 

 

version

message is used when the counter cannot detect a

 

 

Numeric data error;

more specific error.

 

 

underflow from con-

 

 

 

version

 

 

 

Numeric data error;

 

 

 

not a number from

 

 

 

conversion

 

 

Error Code -105 to -120 8-3

Page 117
Image 117
Fluke PM6681R, PM6685R manual Error Code -105 to -120