80B/81/85

:STATus :OPERation :CONDition?

Read Operation Status Condition Register

Reads out the contents of the operation status condition register. This register re- flects the state of the measurement process. See figure below. Note that bits 0 to 3, 7, and 9 to 15 are not used.

Returned Format:

<Decimal data> = the sum (between 0 and 368) of all bits that are true. See table below:

Bit No.

Weight

Condition

 

 

 

8

256

Not Measurement

 

 

 

 

6

64

Waiting for bus arming

 

 

 

 

5

32

Waiting for triggering and/ or external arming

 

 

 

 

4

16

Measurement

 

 

 

 

 

 

 

Complies to standards:

SCPI 1991.0, confirmed

Command Reference 9-101

Page 231
Image 231
Fluke PM6681R, PM6685R manual STATus OPERation CONDition?, Read Operation Status Condition Register