:INPut«[1]2» :ATTenuation

￿«<Numeric value>MAXMIN»

Attenuation

PM6680B PM6681

Attenuates the input signal with 1 or 10. The attenuation is automatically set if the input level is set to AUTO.

Parameters:

<Numeric values> ￿ 5, and MIN gives attenuation 1. <Numeric values> > 5, and MAX gives attenuation 10.

Returned format:

1.00000000000E+0001.00000000000E+001 ￿

Example for Input A (1)

SEND:INP:ATT ￿ 10

Example for Input B (2)

SEND￿ :INP2:ATT ￿ 10

*RST condition Input A (1) and Input B (2): 1 (but set by autotrigger since AUTO is on af- ter *RST. (:INP:LEV:AUTO ￿ ON).

Complies to standards:

SCPI 1991.0, confirmed.

 

 

:INPut«[1]2» :COUPling

￿«ACDC»

AC/DC Coupling

PM6680B PM6681

Selects AC coupling (normally used for frequency measurements), or DC cou- pling (normally used for time measurements).

Returned format: ACDC￿

Example for Input A (1)

SEND￿ :INP:COUP ￿ DC

Example for Input B (2)

SEND￿ :INP2:COUP ￿ AC

*RST condition

Input A (1): AC

Input B (2): DC

Complies to standards:

SCPI 1991.0, confirmed.

9-44 Command Reference

Page 174
Image 174
Fluke PM6685R, PM6681R manual INPut«12» ATTenuation, INPut«12» COUPling, AC/DC Coupling