:MEASure :FREQuency :PRF?

[￿ [<exp. val.>[,<res.>]][,<(@«1234567»)>]]

PM6680B/81/85

Pulse Repetition Frequency

Measures the PRF (Pulse Repetition Frequency) of a burst signal.The burst dura- tion must be less than 50% of the pulse repetition frequency (PRF).

￿It is better to set up the measurement with the [:SENS]:FUNC “:FREQ:PRF” command when measuring pulse repetition frequency. This command will allow you to set a suitable sync delay with the [:SENSe]:Sync:DELay command.

How to measure bursts is described in detail in the Operators Manual.

Parameters: <exp. val.> is the expected PRF,

<res.> is the required resolution.

<(@«13456»)> is the channel to measure on: (@1) means input A

(@2) means input B (Only PM6680B and PM6681)

(@3) means input C (HF-input option)

(@4) means input E (Rear panel arming input)

(@5) means input A prescaled by 2

(@6) means the internal reference

(@7) means input A with the variable hysteresis mode (Only PM6680B/81)

If you omit the channel, the instrument measures on input A (@1).

The <expected value> and <resolution> are used to calculate the Measurement Time ([:SENSe]:ACQuisition:APERture). The Sync. Delay is always 10 ￿s (default value)

Complies to standards:

SCPI 1992.0, confirmed.

9-62 Command Reference

Page 192