Fluke PM6681R, PM6685R manual CONFigure ARRay Measuring Function, Command Reference

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 157
Image 157
:CONFigure :ARRay :<Measuring Function>

80B/81/85

:CONFigure :ARRay :<Measuring Function>

￿(<array size>)[,<parameters> [,(<channels>)]]

Configure the counter for an array of measurements

The :CONFigure:ARRay command differs from the :CONFigure command in that it sets up the counter to perform the number of measurements you choose in the <array size>.

To perform the selected function, you must trigger the counter with the :READ:AR- Ray? or :INITiate;:FETCh:ARRay? queries.

Parameters <array size> sets the number of measurements in the array (1 to 2500).

<Measuring Function>, <Parameters>, and <Channels> are defined on page 9-54.

Example:

SEND￿ :CONF:ARR:PER ￿ (7),5E–3,1E–6,(@4)

This example sets up the counter to make seven period measurements. The ex- pected result is 5 ms, and the required resolution is 1 ￿s. The EXT ARM input is the measuring input.

To make the measurements and fetch the seven measurement results:

SEND￿ :READ:ARR? ￿ 7

READ￿ 5.23421E-3,5.12311E-3,5.87526E-3, ￿ 5.50345E-3,5.33901E-3,5.25501E-3, ￿ 5.03571E-3

Complies to standards:

SCPI 1991.0, confirmed.

Command Reference 9-27