:STATus :QUEStionable :ENABle

PM6680B/81/85

:STATus :QUEStionable :ENABle

￿ <Decimal data>

Enable Questionable Data/Signal Status Reporting

Sets the enable bits of the status questionable enable register. This enable register contains a mask value for the bits to be enabled in the status questionable event register. A bit that is set true in the enable register enables the corresponding bit in the status register. See figure on page 9-104.

An enabled bit will set bit #3, QUE (Questionable Status Bit), in the Status Byte Register if the enabled event occurs. See also status reporting on page 3-14.

Power-on will clear this register if power-on clearing is enabled via *PSC.

Parameters:

<dec.data> = the sum (between 0 and 17920) of all bits that are true. See the table on page 9-104.

Returned Format: <Decimal data> ￿

Example:

Send ￿:STAT:QUES:ENAB ￿ 16896

In this example, both ‘unexpected parameter’ bit 14, and ‘overflow’ bit 8, will set the QUE-bit of the Status Byte when a questionable status occurs.

Complies to standards:

SCPI 1991.0, confirmed.

 

 

 

 

:STATus :QUEStionable?

PM6680B/81/85

Read Questionable Data/Signal Event Register

Reads out the contents of the status questionable event register. Reading the Status Questionable Event Register clears the register. See figure on page 9-104.

Returned Format:

<dec.data> = the sum (between 0 and 17920) of all bits that are true. See the table on page 9-104.

Complies to standards:

SCPI 1991.0, confirmed.

Command Reference 9-105