:STATus:OPERation?

80B/81/85

Read Operation Status, Event

Reads out the contents of the operation event status register. Reading the Opera- tion Event Register clears the register. See figure on page 9-101.

Returned Format: <Decimal data>￿

<dec.data> = the sum (between 0 and 368) of all bits that are true. See table on page 9-102.

Complies to standards:

SCPI 1991.0, confirmed.

 

 

 

 

 

 

 

 

:STATus :PRESet

PM6680B/81/85

 

Enable Device Status Reporting

This command has an SCPI standardized effect on the status data structures. The purpose is to precondition these toward reporting only device-dependent status data.

It only affects enable registers. It does not change event and condition registers.

The IEEE-488.2 enable registers, which are handled with the common commands *SRE and *ESE remain unchanged.

The command sets or clears all other enable registers. Those relevant for this counter are as follows:

It sets all bits of the Device status Enable Registers to 1.

It sets all bits of the Questionable Data Status Enable Registers and the Operation Status En- able Registers to 0.

The following registers never change in the counter, but they do conform to the standard

:STATus:PRESet values.

All bits in the positive transition filters of Questionable Data and Operation status registers are 1.

All bits in the negative transition filters of Questionable Data and Operation status registers are 0.

Complies to standards:

SCPI 1991.0, confirmed.