Fluke PM6681R TESTCHECk, TEST SELect, Select Check signal, Select Self-tests, Command Reference

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 244
Image 244
:TEST:CHECk

:TEST:CHECk

￿<Boolean>

Select Check signal

PM6680B/81/85

This command connects the internal reference signal to the measuring logic, in- stead of an external measuring signal. This makes it possible to test all functions.

The frequency of the reference is 10 MHz for PM6680B and PM6685, and 100 MHz for PM6681.

Parameters:

<Boolean> = 1 / ON 0 / OFF

1 and ON means test signal connected

0 and OFF means test signal disconnected.

￿Selecting channel 6 when entering measuring channel for :CONFigure :MEA- Sure, etc., also selects the reference.

Returned format: 10￿

*RST condition: 0

:TEST :SELect

￿«RAM ROM LOGic DISPlay ALL»

PM6680B/81/85

Select Self-tests

Selects which internal self-tests shall be used when self-test is requested by the *TST command.

Returned format:

«RAM ROM LOGic DISPlay ALL»￿

*RST condition: ALL

9-114 Command Reference

Page 244
Image 244
Fluke PM6681R, PM6685R manual TESTCHECk, TEST SELect, Select Check signal, Select Self-tests, Command Reference