:CALCulate :LIMit :LOWer :STATe

:CALCulate :LIMit :LOWer :STATe

￿<Boolean>

Check Against Lower Limit

PM6680B/81/85

Selects if the measured value should be checked against the lower limit.

Parameters <Boolean> = ( 1/ON 0/OFF )

Returned format: 1 0 ￿

*RST condition: 0

Complies to standards:

SCPI 1991.0 confirmed.

 

 

:CALCulate :LIMit :UPPer

￿«<Decimal data>MAXMIN»

PM6680B/81/85

Set Upper Limit

Sets the value of the ‘Upper Limit’, i.e., the highest measurement result allowed before the counter generates a 1 that can be read with :CALCu- late:LIMit:FAIL?, or by reading the corresponding status byte.

Parameters

Range: –9.9*10+37to +9.9*10+37

Returned format:

*RST condition:

<Decimal data>￿

0

Complies to standards:

SCPI 1991.0, confirmed.

9-18 Command Reference