:DISPlay :ENABle

:DISPlay :ENABle

￿< Boolean >

Display State

PM6680B/81/85

Turns On/Off the updating of the entire display section. This can be used for secu- rity reasons or to improve the GPIB speed, since the display does not need to be updated. Turning off the display reduces the dead time between measurements by about 7 ms.

When the display is turned off, the information about the measurement resolution is lost. That is, the counter will always send a full 12 digit mantissa independent of the measurement resolution.

Parameters:

Where <Boolean> = (1 / ON 0 / OFF)

Returned format: 10 ￿

*RST condition: ON

See also: Chapter 6, ‘How to Measure Fast’.

Complies to standards:

SCPI 1991.0, confirmed.