:CALCulate :AVERage :COUNt

￿< No. of samples>

Sample Size for Statistics

Sets the number of samples to use in statistics sampling.

PM6680B PM6681

Parameters: <No. of samples> is a number in the range of 1 to 65535.

Returned format: < No. of samples>￿

*RST condition: 100

:CALCulate :AVERage :STATe

￿< Boolean >

Enable Statistics

PM6680B PM6681

Switches On/Off the statistical function. Note that the CALCulate subsystem is au- tomatically enabled when the statistical functions are switched on. This means that other enabled calculate sub-blocks are indirectly switched on. The statistics must be enabled before the measurements are performed. When the statistical function is enabled, the counter will keep the trigger subsystem initiated until the :CALC:AVER:COUNT variable is reached. This is done without any change in the trigger subsystem settings. Consider that the trigger subsystem is programmed to perform 1000 measurements when initiated. In such a case, the counter must make 10000 measurements if the statistical function requires 9500 measurements because the number of measurements must be a multiple of the number of mea- surements programmed in trigger subsystem (1000 in this example).

Parameters

<Boolean> = ( 1/ON 0/OFF )

Returned format: <10￿

*RST condition: OFF

9-14 Command Reference

Page 144
Image 144
Fluke PM6681R, PM6685R CALCulate AVERage COUNt, CALCulate AVERage STATe, Sample Size for Statistics, Enable Statistics