Fluke PM6681R MEASureMEMoryN?, MEASureMEMory?, Memory Recall, Measure and Fetch Result, Timing

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 188
Image 188
:MEASure:MEMory<N>?

:MEASure:MEMory<N>?

Memory Recall, Measure and Fetch Result

Use this command when you want to measure several parameters fast.

PM6681

:MEAS:MEM1? recalls the contents of memory 1 and reads out the result, :MEAS:MEM2? recalls the contents of memory two and reads out the result etc.

The equivalent command sequence is *RCL1;READ?

The allowed range for <N> is 1 to 9. Use the somewhat slower :MEAS:MEMory? N command described below if you must use memories 10 to 19.

 

TIMING

 

 

 

 

 

 

 

Data Format

 

 

 

 

 

Command

ASCii

 

REAL

 

 

 

 

 

 

:MEAS:MEM1?

7.9 ms

 

6.7 ms

 

 

 

 

 

 

 

 

 

:MEAS:MEM? 1

9.1 ms

 

8.0 ms

 

 

 

 

 

 

 

 

 

*RCL 1;READ?

10.1 ms

 

8.9 ms

 

 

 

 

 

 

 

 

 

Returned format:

 

 

 

 

<measurement result>￿

 

 

Complies to standards:

SCPI 1991.0, confirmed

 

 

 

 

 

 

 

:MEASure:MEMory?

￿<N>

Memory Recall, Measure and Fetch Result

PM6681

Same as above command but somewhat slower. Allows use of all memories (1 to 19).

Example: :MEAS:MEM ￿ 13

This example recalls the instrument setting in memory number 13, makes a meas- urement, and fetches the result.

Complies to standards:

SCPI 1991.0, confirmed

9-58 Command Reference

Page 188
Image 188
Fluke PM6681R, PM6685R MEASureMEMoryN?, MEASureMEMory?, Memory Recall, Measure and Fetch Result, Command Reference, Timing