Fluke PM6685, PM6681R SYSTem PRESet, SYSTem SDETect, Preset, Signal Detection, Command Reference

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 239
Image 239
Preset

PM6680B/81/85

Preset

:SYSTem :PRESet

This command sets the counter to the same default settings as when the front panel key LOCAL/PRESET is pressed in local mode.

￿These are not exactly the same settings as after *RST,

SYST:PRES gives 200 ms Measurement Time and signal detection ON, while *RST gives 10-ms Measurement Time and signal detection OFF.

See also: Default settings on page -.

Complies to standards: SCPI 1991.0, confirmed.

 

:SYSTem :SDETect

PM6685

 

￿ <Boolean>

Signal Detection

This command switches on or off the signal detection, that is, the ability to show NO SIGNAL, NO TRIG on the display.

When signal detection is enabled, the measurement attempt will be abandoned when the no signal is detected. A zero result will be sent to the controller instead of a measurement result, and the timeout bit in the STATus QUEStionable register will be set.

Returned format: «01»￿

Where:

0 means no signal detection

1 means signal detection ON.

*RST condition: 0

Command Reference 9-109

Page 239
Image 239
Fluke PM6681R, PM6685R manual SYSTem PRESet, SYSTem SDETect, Preset, Signal Detection, Command Reference