Fluke PM6681R, PM6685R manual Resolution, Output Subsystem

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 80
Image 80
￿Resolution

Using the Subsystems

This is the value that will determine the output voltage; .00 will give 0 V and .99 will give 5 V. This means that the reading will give:

.67890*5=3.3945 V.

This is ouput as 3.38 V due to the 0.02 V resolution of the analog output.

￿Resolution

The analog output range is 0 to 5 V in 250 steps, so one step is 0.02 V. If the scaling factor is 1, one such step is taken each time the display changes with X.004, and if the scaling factor is 4, one step is taken each time the display changes with X.001.

The X in the above paragraph can be any

digit and does not influence the output

D e fa u lt s c a lin g fa c to r = 1

voltage. If the display changes from 0.996 to 1.000, the voltage drops from 4.98 V to 0V. If the display value in-

creases further, the output voltage starts

Same exponent, opposite sign

 

 

 

 

Output voltage

Scaling factor 1

 

 

 

 

 

4.98 V

 

 

 

 

S c a lin g

fa c to r

=

1 E - 6

 

 

0.004

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20mV

Figure 6-4

To use the shown deci-

 

0.00V

 

 

 

 

 

 

 

 

 

 

2.000 Displayed

 

mal point as reference,

 

0.000

1.000

 

set the exponent of the

 

 

 

0.996

1.996

value

 

 

 

 

 

 

scaling factor to the

 

Output voltage

Scaling factor 4

 

 

same value as the expo-

 

 

 

 

4.98 V

 

 

 

 

 

nent of the measurement

 

 

 

 

 

0.001

 

result but with opposite

 

 

 

 

 

 

 

sign.

 

 

 

 

 

 

20mV

 

 

 

 

 

 

 

 

 

 

 

 

0.00V

 

 

 

 

 

 

 

 

.000

.250

.500

 

Displayed

 

 

 

 

 

 

 

 

 

 

0

 

value

 

 

 

 

.249

0 .499

0

 

 

 

 

 

 

0

0

 

 

 

 

 

 

 

Figure 6-5

Output voltage versus

 

 

 

 

 

 

displayed value for two

 

 

 

 

 

 

different scaling factors.

to increase again; see .

6-12 Output Subsystem

Page 80
Image 80
Fluke PM6681R, PM6685R manual Resolution, Output Subsystem