READ? starts the acquisition and returns the result.

This sequence does the same as the MEA- Sure command, but now it is possible to insert commands between CONFigure and READ? to adjust the setting of a par- ticular function (called fine tuning). For instance, you can set an input attenuator at a required value.

￿CONFigure; INITiate;FETCh?

The READ? command can be divided into the INITiate command, which starts the measurement, and the FETCh? com- mand, which requests the instrument to return the measuring results to the con- troller.

Versatility of Measurement Com- mands

MEASure?

Simple to use,, few

 

additional possibili-

 

ties.

 

 

CONFigure

Somewhat more

READ?

difficult,, but some

 

extra possibilities.

 

 

CONFigure

Most difficult to

INITiate

use,, but many ex-

FETCh?

tra features.

 

 

5-6 MEASurement Function