How to Measure Fast

Block Synchronized Measurements

(:ARM:STARt:LAYer:COUNt 1 and TRIG:COUNt <N>)

Speed Improvement Actions

CAL:INT:AUTO OFF Display OFF

Realtime calculations OFF ([SENS]:INT:FORM PACK)

Low resolution ([SENS]:ACQ:RES LOW)

￿

 

 

￿

￿

 

￿

￿

￿

Dead Time Between Measurements

PM6680B

PM6681

PM6685

9 ms

4.5 ms

9 ms

2 ms

1.3 ms

2.5 ms

0.5 ms

0.12 ms

0.6 ms

0.5 ms

0.025 ms*

0.6 ms

 

 

 

￿1: The GPIB format command will not affect the dead time for the block synchro- nized mode because the counter captures all data before transferring it to the con- troller.

￿2: Switching the real time calculations on/off In the block synchronized mode will significantly decrease the dead time; however, the time for calculations ( 2 ms for PM6680B/85 and 1 ms for PM6681) is added to the transfer time.

*In PM6681, low resolution is used for Back-to-Back period measurements. The measuring time has no effect in this mode. Only the Timestamps are used.

Speed Summary 7-13