PM6680B/81/85

Limit Fail

:CALCulate :LIMit :FAIL?

Returns a 1 if the limit testing has failed (the measurement result has passed the limit), and a 0 if the limit testing has passed.

The following events reset the fail flag:

Power-on

*RST

A :CALC:LIM:STAT OFF ￿ :CALC:LIM:STAT ￿ ON transition

Reading a 1 with this command.

Returned format: 1 0￿

Example:

SEND￿ SENS:FUNC ￿ FREQ;:CALC:LIM:STAT ￿ ON;:CALC:LIM ￿ :HIGH￿ 1E3;READ?;*WAI;:CALC:LIM:FAIL?

READ￿ 1

if frequency ia above 1kHz, otherwize 0

Complies to standards:

SCPI 1991.0, confirmed.

 

 

 

 

PM6680B/81/85

:CALCulate :LIMit :LOWer

￿«<Decimal data>MAXMIN»

Set Low Limit

Sets the value of the ‘Lower Limit’ , i.e., the lowest measurement result allowed be- fore the counter generates a 1 that can be read with :CALCulate:LIMit:FAIL?, or by reading the corresponding status byte.

Parameters

Parameter range: –9.9*10+37to +9.9*10+37.

Returned format: < Decimal data>￿

*RST condition: 0

Complies to standards:

SCPI 1991.0, confirmed.

Command Reference 9-17

Page 147
Image 147
Fluke PM6681R, PM6685R manual CALCulate LIMit FAIL?, CALCulate LIMit LOWer, Limit Fail, Set Low Limit