Autotrigger

PM6680B PM6681

Autotrigger

:INPut :LEVel :AUTO

￿«<Boolean>ONCE»

If set to AUTO, the counter automatically controls both the trigger level and the at- tenuation1. If you have a stable amplitude, use the :AUTO ￿ ONCE, and the autotrigger will determine the trigger level once and then set a fixed level.

￿ From the bus, input A and input B are always set to autotrigger individually.

Parameters:

<Boolean> = ( 1/ON 0/OFF )

ONCE means that the autotrigger switches on, checks the signal, stores the trigger levels as manually set levels, and then switches off auto. This improves measuring speed.

Example for Input A (1)

SEND￿ :INP:LEV:AUTO ￿ OFF

Example for Input B (2)

SEND￿ :INP2:LEV:AUTO ￿ ON

Returned format: 10￿

*RST condition ON

1The autotrigger function normally sets the trigger levels to 50 % of the signal ampli- tude. Two exceptions exists however:

Rise/Fall time measurements: Here the input 1 (A) trigger level is set to 10% and the Input 2 (B) trigger level is set to 90% of the amplitude.

Variable Hysteresis mode (channel 7): The input 1 (A) trigger level is set to 75% and the Input 2 (B) trigger level is set to 25% of the amplitude

Command Reference 9-49