Fluke PM6685 Error Code 1100 to -1139, CNT-8X Device specific errors leading 1 only for PM6681

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 127
Image 127
CNT-8X Device specific errors (leading 1 only for PM6681)

 

 

Error Messages

 

 

 

 

 

 

CNT-8X Device specific errors (leading 1 only for PM6681)

 

 

 

 

 

Error

Error Description

description/explanation/examples

 

Number

 

 

 

(1)100

Device operation

A floating-point error occurred during a counter op-

 

gave floating-point

eration.

 

 

 

 

underflow

 

 

(1)101

Device operation

A floating-point error occurred during a counter op-

 

gave floating-point

eration.

 

 

 

 

overflow

 

 

(1)102

Device operation

A floating-point error occurred during a counter op-

 

gave ‘not a number’

eration.

 

 

 

 

 

 

 

(1)110

Invalid measure-

The counter was requested to set a measurement

 

ment function

function it could not make.

 

 

 

(1)120

Save/recall memory

An attempt was made to write in a protected mem-

 

protected

ory.

 

 

 

(1)130

Unsupported com-

Indicates a mismatch between bus and counter ca-

 

mand

pabilities.

 

 

 

(1)131

Unsupported

 

 

boolean command

 

 

 

 

 

(1)132

Unsupported deci-

 

 

mal command

 

 

 

 

 

(1)133

Unsupported enu-

 

 

merated command

 

 

 

 

 

(1)134

Unsupported auto

 

 

command

 

 

 

 

 

(1)135

Unsupported single

 

 

shot command

 

 

 

 

 

(1)136

Command queue

The counter has an internal command queue with

 

full; last command

room for about 100 commands. A large number of

 

 

 

 

discarded

commands arrived fast without any intervening

 

 

 

query.

 

(1)137

Inappropriate suffix

A suffix unit was not appropriate for the command.

 

unit

Recognized units are Hz (Hertz), s (seconds), Ohm

 

 

 

 

 

() and V (Volt).

 

(1)138

Unexpected com-

A command reached counter execution which

 

mand to device exe-

should have been handled by the bus.

 

 

 

 

cution

 

 

(1)139

Unexpected query

A query reached counter execution which should

 

to device execution

have been handled by the bus.

 

 

 

Error Code (1)100 to -(1)139 8-13

Page 127
Image 127
Fluke PM6685 Error Code 1100 to -1139, CNT-8X Device specific errors leading 1 only for PM6681, Error Description, Number