Fluke PM6685R, PM6681R TRIGgerCOUNt, No. of Triggerings on each Ext Arm start, Command Reference

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 246
Image 246
:TRIGger:COUNt

:TRIGger:COUNt

￿«<Numeric value> MIN MAX»

PM6680B/81/85

No. of Triggerings on each Ext Arm start

Sets how many measurements the instrument should make for each ARM:STARt condition, (block arming).

These measurements are done without any additional arming conditions before the measurement. This also means that stop arming is disabled for the measure- ments inside a block.

￿The actual number of measurements made on each INIT equals to:

(:ARM:START:COUN)*(:TRIG:START:COUNT)

Parameters:

<Numeric value> is a number between 1 and 65535. MAX gives 65535

MIN gives 1

Example:

SEND￿ :TRIG:COUN ￿ 50

Returned format: <Numeric value>￿

*RST condition: 1

Complies to standards:

SCPI 1991.0, confirmed.

9-116 Command Reference