EXPLANATIONS OF THE MEASURING FUNCTIONS

This sub-chapter explains the various measurements that can be done with :MEASure and :CONFigure;:READ. Only the queries for single measurements using the measure command are given here, but all of the information is also valid for the :CONFigure command and for both scalar (single) and array measurements.

 

:MEASure_«:DCYCle/:PDUTycycle»

PM6680B/81/85

 

[￿ [<threshold>] [,(@«1246»)]]

Positive Duty Cycle

Traditional duty cycle measurement is performed. That is, the ratio between the on time and the off time of the input pulse is measured.

Parameters

<threshold> parameter sets the trigger levels in volts. If omitted, the auto trigger level is set to 50 percent of the signal.

(@«1246») is the channel to measure on: (@1) means input A

(@2) means input B (Only PM6680B and PM6681)

(@4) means input E (Rear panel arming input)

(@6) means the internal reference

If you omit the channel, the instrument measures on input A (@1).

Example:

SEND￿ :MEAS:PDUT?

READ￿ +5.097555E-001

In this example, the duty cycle is 50.97%

Complies to standards:

SCPI 1991.0, confirmed.

Command Reference 9-59

Page 189
Image 189
Fluke PM6681R, PM6685R manual MEASure«DCYCle/PDUTycycle», Positive Duty Cycle, Send MEASPDUT?, Read +5.097555E-001