:FETCh :ARRay?

PM6680B/81/85

:FETCh :ARRay?

￿«<fetch array size>MAX»

Fetch an Array of Results

:FETCh:ARRay? query differs from the :FETCh? query by fetching several mea- suring results at once.

An array of measurements must first be made by the commands. :INITiate, :MEASure:ARRay? or :CONFigure:ARRay;:READ?

If the array size is set to a positive value, the first measurement made is the first result to be fetched.

When the counter has made an array of measurements, :FETCh:ARRay? ￿ 10 fetches the first 10 measuring results from the output queue. The second :FETCh:ARRay? ￿ 10 fetches the result 11 to 20, and so on. When the last mea- suring result has been fetched, fetch:array starts over again with the first result.

In totalizing for instance, you may want to read the last measurement result in- stead of the first one. This is possible if you set the array size to a negative num- ber. Example: :FETCh:ARRay? ￿ –5 fetches the last five results. The output queue pointer is not altered when the array size is negative. That is, the example above always gives the last five results every time the command is sent.

:FETCh:ARRay? ￿ –1 is useful to fetch intermediate results in free-running or ar- ray measurements without interrupting the measurement.

Parameters

:ARRay means that an array of retrievals are done for each :FETCh command. <fetch array size> is the number of retrievals in the array. This number must not exceed the number of measuring results in the measurement result buffer. The <SIZE> parameter maximum limit is depending on the :SENSe:INTernal:FORMat command as follows:

 

 

Array Size

 

 

 

 

Format
Measuring function
PM6680B/85
PM6681

 

 

 

 

 

 

Real:
All functions

2048

7019

 

 

 

 

 

Packed:
Frequency, Period, Ratio Totalize

2166

6143

 

 

 

 

 

 

Pulse Width

764

4466

 

 

 

 

 

 

 

Time-Interval, Rise/Fall time

 

4466

 

 

 

 

 

 

 

 

Phase, Duty Cycle,Volt

 

7019

 

 

 

 

 

 

 

 

Low resolution Frequency and Period

 

8191

 

 

 

 

 

 

 

 

Low Res. Time-Interval and Pulse Width

 

4095

 

 

 

 

 

 

 

 

 

 

 

MAX means that all the results in the output buffer will be fetched.