Fluke PM6681R, PM6685R manual Speed Summary

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 109
Image 109
Speed Summary

How to Measure Fast

Speed Summary

The following table summarizes the time that can be gained when fine tuning the measurement process.

The normal dead time between frequency measurements is approximately as fol- lows:

48.8ms for the PM6680B.

85 ms for the PM6681.

75 ms for the PM6685.

If you should read out the measuring re- sult to the controller add read out time to each result. (Consult your controller man- ual.)

Individually Synchronized Measurements

Speed Improvement

 

 

 

 

 

 

 

 

Actions

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

None, normal READ?

AUTOtrigger OFF

INT:CONT and GET

Display OFF

CAL:INT:AUTO OFF

100ns resolution

Dead Time Between Measurements Including

 

Transfer to Controller

 

 

 

 

 

 

 

 

ASCII Data Format

Real Data Format

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PM6680B

PM6681

PM6685

PM6680B

PM6681

PM6685

 

 

 

 

 

 

 

 

 

 

 

 

 

￿

 

 

 

 

 

Approx.

Approx.

Approx.

Approx.

 

Approx.

Approx.

 

 

 

 

 

50 ms

85 ms*

75 ms

45 ms

 

85 ms*

75 ms

 

 

 

 

 

 

 

 

￿

 

 

 

 

19 ms

10 ms

23 ms

15 ms

 

9 ms

20 ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

￿

￿

 

 

 

12 ms

9 ms

20 ms

9 ms

 

8 ms

17 ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

￿

￿

￿

￿

 

12 ms

5.5 ms

12 ms

9 ms

 

4.2 ms

9 ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

￿

￿

￿

￿

￿

12 ms

5.5 ms

12 ms

8 ms

 

4.2 ms

8 ms

 

 

 

 

 

 

 

 

 

 

 

 

 

￿Turning the real-time calculations on/off will not affect the dead time because the calculations are still done inside the measurement loop during the output of data.

*It takes longer time for PM6681 to determine trigger levels than for PM6680B, why? The reason is that PM6681 must find the correct level from 16 times as many triggerlevel setting steps than thePM6680B/85 (1.25 mV steps versus 20mV steps).

Speed Summary 7-11

Page 109
Image 109
Fluke PM6681R, PM6685R manual Speed Summary