:INPut :FILTer

PM6680B/81/85

 

￿ <Boolean>

Low Pass Filter

Switches on or off the low pass filter on input 1 (A). It has a cutoff frequency of 100 kHz.

Parameters:

<Boolean> is (1 / ON 0 / OFF)

Returned format: 10￿

*RST condition OFF

Complies to standards:

SCPI 1991.0, confirmed.

 

 

 

 

PM6685

:INPut :HYSTeresis

￿«<Decimal data>MAX MIN»

Sensitivity

The sensitivity setting on the front panel is called HYSTeresis from the bus. The range is 27.12 mV to 75.4 V. This setting has no effect unless autosensitivity is turned off, see the following page.

Note that the sensitivity setting is coupled with the hysteresis setting according to the formula:

Trigger Level ￿ Hysteresis ￿ 37.7057 ￿

2

Parameters: <Decimal data> is a number between 27.12E-3 and 75.4.

MAX gives +75.4 V, MIN gives +2.7 mV

When using MAX as data, the counter always tries to set the hysteresis to +75.4 V. Unless the Trigger level is set to 0, this setting is impossible, and the counter will return an error message.

Returned format: <Decimal data>

Example:

SEND:INP:HYST ￿ 0.5;HYST:AUTO ￿ 0

This example sets the sensitivity to 0.5 V and switches off autosensitivity. *RST condition 0.65 V (but controlled by Autotrigger since AUTO is on after *RST)

Command Reference 9-45

Page 175
Image 175