:MEASure :<Measuring Function>?

:MEASure :<Measuring Function>?

[￿ [<parameters>][ ,(<channels>)]]

PM6680B/81/85

Make one measurement

The measure query makes a complete measurement, including configuration and readout of data. Use measure when you can accept the generic measurement without fine tuning.

￿When a CONFigure command or MEASure? query is issued, all counter settings are set to the *RST settings., except those specified as <parameters> and <channels> in the CONFigure command or MEASure? query.

You cannot use the :MEASure? query for :TOTalize:CONTinuous, since this function measures without stopping (continuously forever).

The :MEASure? query is a compound query identical to: :ABORt;:CONFigure:<Meas_func>;:READ?

Parameters:

<Measuring Function>, <Parameters> and <Channels> are defined on page 9-54.You may omit <parameters> and <Channels>, which are then set to default.

Returned format: <data>￿

Where: The format of the returned data is determined by the format commands: :FORMat and :FORMat:FIXed.

Example:

SEND￿ :MEAS:FREQ? ￿ (@3)

READ￿ 1.78112526833E+009

This example measures the frequency on the C-input and outputs the result to the controller.

Type of command: Aborts all previous measurement commands if *WAI is not used.

See also: ‘Explanations of the Measuring Functions’ starting on page 9-59.

Complies to standards:

SCPI 1991.0, confirmed.

9-56 Command Reference