:ARM :STOP :SLOPe

PM6680B/81/85

 

￿ «POSitive NEGative»

External Stop Arming Slope

Sets the slope for the stop arming condition.

Returned format: POSNEG￿

Example:

SEND￿ :ARM:STOP:SLOP ￿ NEG￿

*RST condition: POS

Complies to standards:

SCPI 1991.0, confirmed.

 

 

 

 

PM6680B/81/85

:ARM :STOP :SOURce

￿«EXTernal2 EXTernal4 IMMediate»

External Stop Arming Source

Selects between channel 2 (Input B) and channel 4 (Input E) as stop arming input, or switches off the stop arming function.

Parameters:

 

EXTernal2 is input B

(Only PM6680B/81)

EXTernal4 is input E

 

IMMediate is Stop arming OFF

Returned format: EXT4IMM￿

Example:

SEND￿ :ARM:STOP:SOUR ￿ EXT4￿

*RST condition: IMM

Complies to standards:

SCPI 1991.0, confirmed.

Command Reference 9-11

Page 141
Image 141
Fluke PM6681R, PM6685R manual ARM Stop SLOPe, ARM Stop SOURce, External Stop Arming Slope, External Stop Arming Source