Gate On/Off

PM6680B/81/85

Gate On/Off

:TOTalize :GATE

￿ <Boolean>

Open/closes the gate for :TOTalize[:CONTinuous].

￿Before opening the gate with this command, the counter must be in the ‘contin- uously initiated’ state , (:INIT:CONT ￿ ON)or else the totalizing will not start.

Parameters: <Boolean> = (1 / ON 0 / OFF)

Returned format:

<Boolean>￿

 

 

Example:

 

 

 

 

Send ￿ :FUNC

￿

‘TOT ￿

1’

Selects totalizing

on input

A

 

 

Send ￿ :INIT:CONT ￿ ON;TOT:GATE ￿ ON

 

This will initiate totalizing, reset the totalizing value to zero, and start totalizing.

Send ￿ TOT:GATE

￿ OFF

 

Stop totalizing

Read ￿ :FETCh:ARRay? ￿

-1

Read the final result

*RST condition:

OFF

 

 

 

:VOLTage:GATed:STATe

PM6680B PM6681

 

￿ <Boolean>

Gated Voltage Measurement

Selects the gated mode for the :VOLTage:MAXMINPTPeak measuring func- tions and for the Autotrigger function.

The gated mode is useful for removing overshoot and undershoot. The gate signal is controlled by the :ARM:STOP:SLOPe and :ARM:STOP:SOURce commands. If channel 2 (B) is the source for the gating signal, all other characteristics of that channel can be used. When Gated Voltage is selected, the Stop Arming function is disabled from its normal stop arming usage. When gated voltage mode is selected, high enables measurement and low disables measurements. Use the slope if you want it the other way around.

Parameters:

<Boolean> = 1/ON0/OFF

Returned format: 10 ￿

*RST condition: OFF

Command Reference 9-97