Fluke PM6681R, PM6685R SYSTem TOUT TIME, Timeout On/Off, Timeout, Set, Command Reference

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 241
Image 241
:SYSTem :TOUT
Returned format:

 

:SYSTem :TOUT

PM6680B/81/85

 

￿ <Boolean>

Timeout On/Off

This command switches on or off the timeout. When timeout is enabled, the mea- surement attempt will be abandoned when the time set with :SYST:TOUT:TIME has elapsed. A zero result will be sent to the controller instead of a measurement result and the timeout bit in the STATus QUEStionable register will be set.

0 means no timeout; 1 means that the timeout set by :SYS- Tem:TOUT:TIME is used.

Example:

SEND￿ :SYST:TOUT ￿ 1;TOUT:TIME ￿ 0.5;:STAT:QUES:ENAB ￿ 1024;:*SRE ￿ 8

This example turns on timeout, sets the timeout to 0.5 s, enables status reporting of questionable data at timeout, and enables service request on questionable data.

SEND￿ *STB?

If bit 3 in the status byte is set, read the

questionable data

status.

SEND￿ :STAT:QUES:EVEN?

This query reads the ques-

tionable

data status.

READ￿ «10240»

 

1024 means timeout has occurred, and 0

means no

timeout.

 

*RST condition: 0

PM6680B/81/85

:SYSTem :TOUT :TIME

￿«<Numeric value>MINMAX»

Timeout, Set

This command sets the timeout in seconds.

The timeout starts when a measurement starts and, if no result is obtained when the set timeout has elapsed the measurement is terminated.

Note that you must enable timeout using :SYST:TOUT_ON for this setting to take effect.

Parameters:

<Numeric value> is the timeout in seconds. The range is 0.1 to 25.5 s for PM6680B and PM6685. The range is 64 ms to 400 s for PM6681

MIN gives 0.1 s (64 ms for PM6681)

MAX gives 25.5 s (400 s for PM6681)

Returned format:

*RST condition:

<Numeric value>￿

0.1 (6.4E-2 for PM6681)

Complies to standards:

SCPI 1991.0, confirmed.

Command Reference 9-111