Fluke manual Error Code 1214 to -1240, CNT-8X Device specific errors leading 1 only for PM6681

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 129
Image 129
Error Code (1)214 to -(1)240 8-15

 

 

Error Messages

 

 

 

 

 

 

CNT-8X Device specific errors (leading 1 only for PM6681)

 

 

 

 

 

Error

Error Description

description/explanation/examples

 

Number

 

 

 

(1)214

Hash table overflow

The hash table was too small to hold all mnemon-

 

 

ics. Ordinarily indicates a failure to read (RAM or

 

 

 

 

 

 

ROM) correctly. Contact your local service center.

 

(1)220

Parser error

Generic error in the parser.

 

 

 

 

 

 

 

 

(1)221

Illegal parser call

The parser was called when it should not be active.

 

 

 

 

 

 

 

 

(1)222

Unrecognized input

A character not in the valid IEEE488.2 character set

 

character

was part of a command.

 

 

 

(1)223

Internal parser error

The parser reached an unexpected internal state.

 

 

 

 

 

 

 

 

(1)230

Response formatter

Generic error in the response formatter.

 

error

 

 

 

 

 

(1)231

Bad response for-

The response formatter was called when it should

 

matter call

not be active.

 

 

 

(1)232

Bad response for-

The response formatter was called to output an end

 

matter call (eom)

of message, when it should not be active.

 

 

 

(1)233

Invalid function

The response formatter was requested to output

 

code to response

data for an unrecognized function.

 

 

 

 

formatter

 

 

(1)234

Invalid header type

The response formatter was called with bad data for

 

to response format-

the response header (normally empty)

 

 

 

 

ter

 

 

(1)235

Invalid data type to

The response formatter was called with bad data for

 

response formatter

the response data.

 

 

 

(1)240

Unrecognized error

An error number was found in the error queue for

 

number in error

which no matching error information was found.

 

 

 

 

queue

 

 

See also Error Messages in Appendix 1 of the Operators Manual.