:MEASure :TOTalize :GATed?

[￿ (@«12456») [,(@«12456»)]]

PM6680B PM6681

Totalize X gated by Y

The counter totalizes the pulses on the primary channel. The totalizing starts when the gate signal on the secondary channel goes on and stops when the gate signal goes to off. The polarity of on/off is controlled via the :INPut:SLOPe command of the gate signal.

Select if the counter should count positive or negative transitions with the :INPut:SLOPe command of the measuring channel.

Parameters

The first <(@«12456»)> is the channel to measure on, the second one is the gate channel: (@1) means input A

(@2) means input B

(@4) means input E (rear panel arming input)

(@5) means input A prescaled by 2

(@6) means the internal reference

If you omit the channels, the instrument measures on input A with input B as the gate channel.

:MEASure :TOTalize :SSTop?

[￿ ,(@«12456») [,(@«12456»)]]

PM6680B PM6681

Totalize X start/stop by Y

The counter totalizes the pulses on the primary channel. The totalizing starts when the gate signal on the secondary channel goes on and stops the next time the gate signal goes on. The polarity of ON is controlled via the :INPut:SLOPe command of the Start /stop channel.

Select if the counter should count positive or negative transitions with the :IN- Put:SLOPe command of the measuring channel.

Parameters:

The first <(@«12456»)> is the channel to measure on, and the second one is the start/stop channel:

(@1) means input A

(@2) means input B

(@4) means input E (rear panel arming input)

(@5) means input A prescaled by 2

(@6) means the internal reference

If you omit the channels, the instrument measures on input A with input B is the start/stop channel.

9-72 Command Reference

Page 202
Image 202
Fluke PM6685R, PM6681R MEASure TOTalize GATed?, MEASure TOTalize SSTop?, Totalize X gated by Y, Totalize X start/stop by Y