PM6680B/81/85

*ESE

￿<Decimal data>

Standard Event Status Enable

Sets the enable bits of the standard event enable register. This enable register contains a mask value for the bits to be enabled in the standard event status regis- ter. A bit that is set true in the enable register enables the corresponding bit in the status register. An enabled bit will set the ESB (Event Status Bit) in the Status Byte Register if the enabled event occurs. See also status reporting on page 3-14.

Parameters: <dec.data> = the sum (between 0 and 255) of all bits that are true.

Event Status Enable Register (1 = enable)

Bit

7

6

5

4

3

2

1

0

Weight

Enables

128

PON, Power-on occurred

64

URQ, User Request

32

CME, Command Error

16

EXE, Execution Error

8

DDE, Device Dependent Error

4

QYE, Query Error

2

RQC, Request Control (not used)

1

Operation Complete

Returned Format: <Decimal data> ￿

Example:

SEND￿ *ESE ￿ 36

In this example, command error, bit 5, and query error, bit 2, will set the ESB-bit of the Status Byte if these errors occur.

Figure 9-3Bits in the standard event status register.

Complies to standards:

IEEE 488.2 1987.

Command Reference 9-121

Page 251
Image 251
Fluke PM6681R, PM6685R manual Standard Event Status Enable, Event Status Enable Register 1 = enable, Operation Complete