:ACQuisition :HOFF

￿<boolean>

Hold Off On/Off

Switches the Hold Off function On/Off.

Parameters:

<Boolean> = 1 / ON 0 / OFF

Returned format: 1 0￿

*RST condition: OFF

PM6680B PM6681

:ACQuisition :HOFF: ECOunt

￿«<Decimal value>MINMAX»

Hold Off, set event counter

PM6680B PM6681

Sets the Hold Off event value. The counter counts negative events on the B input (channel 2).

Parameters:

<decimal value> is a number in the range 2 to 16 777 215.

Returned format: <Decimal value>￿

*RST condition: 100

9-88 Command Reference

Page 218
Image 218
Fluke PM6685R, PM6681R manual ACQuisition Hoff ECOunt, Hold Off On/Off, Hold Off, set event counter