Fluke PM6685, PM6681R ACQuisition HOFF MODE, ACQuisition HOFF TIME, Hold Off Mode, Hold Off Time

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 219
Image 219
Hold Off Mode

PM6680B PM6681

Hold Off Mode

:ACQuisition :HOFF :MODE

￿ «TIMEEVENt»

Selects if triggering is going to be disabled for a preset time or for a preset number of events.

When set to event, the counter counts negative edges on the B input (channel 2).

This function is coupled to the :ARM:START:DEL, :ARM:START:ECO, :ARM:STOP:DEL and :ARM:STOP:ECO. The different delays must all be of the same type, (Time or Event). This means that setting one of them to Event delay causes the others to be set to Event delays.

Parameters:

TIME

EVENt

Returned format:

TIME EVENt￿

*RST condition:

TIME

PM6680B PM6681

Hold Off Time

:ACQuisition :HOFF :TIME

￿«<Decimal value> MINMAX»

Sets the Hold Off time value.

Parameters:

<Decimal data> = a number between 200E–9 and 1.6777215 for PM6680B, and between 40E–9 and 1.34217727 for PM6681.

Returned format:

<Decimal value>￿

*RST condition:

10 ￿s for PM6680B and 1 ￿s for PM6681

Command Reference 9-89

Page 219
Image 219
Fluke PM6685, PM6681R manual ACQuisition HOFF MODE, ACQuisition HOFF TIME, Hold Off Mode, Hold Off Time, Command Reference