PM6680B/81/85

:MEASure :PERiod?

[￿ [<expected value> [,<resolution>]][,<(@«1234567»)>]]

Period

A traditional period measurement is performed.

The <expected value> and <resolution> are used to calculate the Measurement Time ([:SENSe]:ACQuisition:APERture).

Parameters:

<expected value> is the expected Period,

<resolution> is the required resolution,

<(@«123456»)> is the channel to measure on: (@1) means input A

(@2) means input B (Only PM6680B and PM6681)

(@3) means input C (HF-input option)

(@4) means input E (Rear panel arming input)

(@5) means input A prescaled by 2

(@6) means the internal reference

(@7) means input A with the variable hysteresis mode (Only PM6680B)PM6681)

If you omit the channel, the instrument measures on input A (@1).

Complies to standards:

SCPI 1991.0, confirmed.

 

 

PM6680B PM6681

Phase

:MEASure :PHASe?

[￿ [<expected value>[,<resolution>]] [,(@«12»),(@«12»)]]

A traditional PHASe measurement is performed.

Parameters:

<expected value> and <resolution> are ignored by the counter

The first (@«12») is the start channel and the second (@«12») is the stop channel (@1) means input A

(@2) means input B

If you omit the channel, the instrument measures between input A and input B.

Complies to standards:

SCPI 1991-0, approved.