*OPC?

PM6680B/81/85

Operation Complete Query

Operation Complete query. The Operation Complete query places an ASCii char- acter 1 into the device’s Output Queue when all pending selected device opera- tions have been finished.

Returned Format: 1￿

See also:

Example 6 is Chapter 4.

Complies to standards:

IEEE 488.2 1987.

 

 

 

 

 

 

 

 

*OPT?

PM6680B/81/85

 

Option Identification

Response is a list of all detectable options present in the instrument, with absent options represented with an ASCii ‘0’.

Returned format:

<Bus option>,<Prescaler option>￿

Where:

<Bus option> = GPIB <Prescaler option> = 01020

0 for prescaler option means that no prescaler is installed.

￿Oscillator type are not detectable and can therefore, not be reported.

Complies to standards:

IEEE 488.2 1987.