Using the Subsystems

￿Example:

SENDCONFigure:FREQ￿2E6,1

2E6 is the expected value

1 is the required resolution (1Hz)

SENDINPut:IMPedance50￿OHM

Sets input impedance to 50 Ω

SENDREAD?

Starts the measurement and returns the result.

CONFigure;INITiate;FETCh?

The READ? command can be divided into the INITiate command, which starts the measurement, and the FETCh? com- mand, which requests the instrument to return the measuring results to the con- troller.

￿Example:

SENDCONFigure:FREQ￿20E6,1

20E6 is the expected signal value

1 is the required resolution

SENDINPut:IMPedance1E￿6

Sets input impedance to 1 MΩ

SENDINITiate

Starts measurement

SENDFETCh?

Fetches the result

Versatility of measurement com-

mands

MEASure?

Simple to use, few addi-

 

tional possibilities.

CONFigure

Somewhat more difficult,

READ?

but some extra possibili-

 

ties.

CONFigure

Most difficult to use, but

INITiate

many extra features.

FETCh?

 