Using the Subsystems

Trigger/Arming Subsystem

The SCPI TRIGger subsystem enables syn- chronization of instrument actions with specified internal or external events. The following list gives some examples.

Instrument Action

Some examples of events to synchronize with are as follows:

measurement

bus trigger

external signal level or pulse

10 occurrences of a pulse on the external trigger input

other instrument ready

signal switching

input signal present

1 second after input signal is present

sourcing output signal

switching system ready

The ARM-TRIG Trigger Configuration

Figure 6-15gives a typical trigger config- uration, the ARM-TRIG model. The con- figuration contains two event-detection layers: the ‘Wait for ARM’ and ‘Wait for TRIG’ states.

 

 

 

 

 

 

 

 

 

 

Default state

ABORt

 

 

Idle

 

 

 

 

 

 

 

 

 

 

 

 

*RST

 

 

 

 

after power-on

 

 

 

 

 

 

 

 

 

 

 

 

 

or reset

 

pon

 

 

 

 

 

 

 

 

 

 

Trigger system initiated

 

No longer

 

initiated

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Trigger system

 

 

 

 

 

 

Initiated

 

initiated

 

Still initiated

 

Completed No.

 

 

 

 

 

 

of ARM loops

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Arm Layer

 

 

 

 

 

Wait for ARM

 

ARM conditions

 

Completed

 

No. of TRIGger

 

 

satisfied

 

 

 

 

loops

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Trigger Layer

 

 

 

 

 

Wait for TRIG

 

 

 

 

 

 

 

 

TRIGger conditions

 

Instrument

 

 

satisfied

 

Actions

 

 

 

 

 

 

 

 

complete

 

 

 

 

 

 

 

 

 

 

Instrument

Actions

Figure 6-15Generalized ARM-TRIG model.

Trigger/Arming Subsystem 6-27