Fluke PM6681R, PM6685R Error Code -200 to -221, Execution errors, Error Description, Number

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 121
Image 121
Execution errors

 

 

Error Messages

 

 

 

 

 

 

 

Execution errors

 

 

 

 

 

Error

Error Description

description/explanation/examples

 

Number

 

 

 

–200

Execution error

This is the generic syntax error for devices that can-

 

 

not detect more specific errors. This code indicates

 

 

 

 

 

 

only that an Execution Error as defined in

 

 

 

IEEE-488.2, 11.5.1.1.5 has occurred.

 

–210

Trigger error

 

 

 

 

 

 

 

 

 

–211

Trigger ignored

Indicates that a GET, TRG, or triggering signal was

 

 

 

received and recognized by the counter but was ig-

 

 

 

nored because of counter timing considerations; for

 

 

 

example, the counter was not ready to respond.

 

–212

Arm ignored

Indicates that an arming signal was received and

 

 

recognized by the counter but was ignored.

 

 

 

 

–213

Init ignored

Indicates that a request for a measurement initiation

 

 

was ignored because another measurement was al-

 

 

 

 

 

 

ready in progress.

 

–214

Trigger deadlock

Indicates that the trigger source for the initiation of a

 

 

measurement is set to GET and subsequent mea-

 

 

 

 

 

 

surement query is received. The measurement can-

 

 

 

not be started until a GET is received, but the GET

 

 

 

would cause an INTERRUPTED error.

 

–215

Arm deadlock

Indicates that the arm source for the initiation of a

 

 

measurement is set to GET and subsequent mea-

 

 

 

 

 

 

surement query is received. The measurement can-

 

 

 

not be started until a GET is received, but the GET

 

 

 

would cause an INTERRUPTED error.

 

–220

Parameter error

Indicates that a program-data-element related error

 

 

occurred. This error message is used when the

 

 

 

 

 

 

counter cannot detect the more specific errors –221

 

 

 

to –229.

 

–221

Settings conflict

Indicates that a legal program data element was

 

Settings conflict;

parsed but could not be executed due to the current

 

 

 

 

PUD memory is

counter state (see IEEE-488.2, 6.4.5.3 and

 

 

protected

11.5.1.1.5.)

 

 

Settings conflict; in-

 

 

 

valid combination of

 

 

 

channel and function

 

 

Error Code -200 to -221 8-7