Fluke PM6681R, PM6685R manual MEASure TOTalize ACCumulated?, Totalize X gated by Y, accumulated

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 200
Image 200
:MEASure :TOTalize :ACCumulated?

:MEASure :TOTalize :ACCumulated?

[￿ <time for gate open>][,(@«12456») [,(@«12456»)]]

PM6680B PM6681

Totalize X gated by Y, accumulated

The counter totalizes the pulses on the primary channel. The totalizing starts when the gate signal on the secondary channel goes on and stops when the gate signal goes to off. The polarity of on/off is controlled via the :INPut:SLOPe command of the gate channel. The result is the sum of counts in all the gate openings that oc- cur during a preset time <time for gate open>.

If you use the :CONFigure command, you can select if the counter should count positive or negative transitions with the :INPut:SLOPe command of the measur- ing channel.

Parameters: <time for gate open> is the time you want the totalizing to proceed. Range PM6680B: is 0.8E–6, 1.6E–6, 3.2E–6, 6.4E–6, 12.8E–6, and 50E–6 to 400 s Range PM6681 and 80E–9, 160E–9, 320E–9, 640E–9, 1.28E–6, and 20E–6 to 400 s.

The first <(@«12456»)> is the channel to measure on.

The second <(@«12456»)> is the gate channel. (@1) means input A
(@2) means input B

(@4) means input E (rear panel arming input)

(@5) means input A prescaled by 2

(@6) means the internal reference

￿If you omit the channels, the instrument measures on input A with input B as the gate channel.

*RST condition:

Time for gate open = 10 ms ([:SENSe]ACQuisition:APERture)

9-70 Command Reference

Page 200
Image 200
Fluke PM6681R, PM6685R manual MEASure TOTalize ACCumulated?, Totalize X gated by Y, accumulated, Command Reference