Fluke PM6685R OUTPut SCALe, Enable Analog Out, Scaling Factor, Analog Output, Scaling factor

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 210
Image 210
:OUTPut

:OUTPut

￿<Boolean>

PM6680B/81/85

Enable Analog Out

This command switches on/off the analog output. See also :OUTput:SCALe com- mand on the next page.

Parameters

<Boolean> = ( 1/ON 0/OFF )

Returned format: <10>￿

Example:

Send￿ :OUTP ￿ 1

Switches on the analog output.

*RST condition: OFF

Complies to standards:

SCPI 1991.0 confirmed.

 

 

:OUTPut :SCALe

￿< Decimal data >

Scaling Factor, Analog Output

PM6680B/81/85

This command sets the scaling factor for the analog output. The measurement re- sult is scaled after math, if math is used.

If you want a full-scale output for a specific readout, the formula is:

Scaling factor ￿

1

full scale value

Parameters

<Decimal data> is the scaling factor. The range is –1020 to +1020.

Returned format: < Decimal data>￿

Example:

If you want full scale output (5 V) for a reading of 0.00359, Scaling factor ￿ 0.00359 ￿ 0.000718

5

Send￿ :OUTP:SCAL ￿ 718E–6

*RST condition: 1

9-80 Command Reference

Page 210
Image 210
Fluke PM6685R manual OUTPut SCALe, Enable Analog Out, Scaling Factor, Analog Output, Command Reference, Scaling factor