Fluke PM6681R manual STATus OPERation ENABle, Enable Operation Status Reporting, Command Reference

Models: PM6685 PM6685R PM6681 PM6681R

1 276
Download 276 pages 1.51 Kb
Page 232
Image 232
:STATus :OPERation :ENABle

:STATus :OPERation :ENABle

￿<Decimal data>

80B/81/85

Enable Operation Status Reporting

Sets the enable bits of the operation status enable register. This enable register contains a mask value for the bits to be enabled in the operation status event reg- ister. A bit that is set true in the enable register enables the corresponding bit in the status register. See figure on page 9-101.

An enabled bit will set bit #7, OPR (Operation Status Bit), in the Status Byte Regis- ter if the enabled event occurs. See also status reporting on page 3-14.

Power-on will clear this register if power-on clearing is enabled via *PSC.

Parameters: <dec.data> = the sum (between 0 and 368) of all bits that are true. See ta- ble below:

Bit No.

Weight

 

Condition

 

 

 

 

8

256

 

No measurement

 

 

 

 

 

 

6

64

 

Waiting for bus arming

 

 

 

 

 

 

5

32

 

Waiting for triggering and/or external arming

 

 

 

 

 

 

4

16

 

Measurement

 

 

 

 

 

 

 

 

 

Returned Format:

<Decimal data>￿

Example:

 

 

 

SEND￿ :STAT:OPER:ENAB

￿ 288

In this example, waiting for triggering, bit 5, and Measurement stopped, bit 8, will set the OPR-bit of the Status Byte. (This method is faster than using *OPC if you want to know when the measurement is ready.)

Complies to standards:

SCPI 1991.0, confirmed.

9-102 Command Reference

Page 232
Image 232
Fluke PM6681R STATus OPERation ENABle, Enable Operation Status Reporting, Command Reference, Bit No, Weight, Condition