4 Operation | Model 3000PA | |
|
|
|
|
|
|
|
|
|
operation, such as setting the instrument span or zero setting, adjusting the alarm setpoints, or defining analysis ranges.
After a password is assigned, the operator must log out to activate it. Until then, anyone can continue to operate the instrument without entering the new password.
Only one password can be defined. Before a unique password is assigned, the system assigns TBEAI by default. This allows access to anyone. After a unique password is assigned, to defeat the security, the password must be changed back to TBEAI.
•Logout: Logging out prevents an unauthorized tampering with analyzer settings.
•More: Select and enter More to get a new screen with additional subfunctions listed.
•
•Version: Displays Manufacturer, Model, and Software Version of instrument.
4.3.1 Setting the Display
Contrast Function is DISABLED
(Refer to Section 1.6)
If you cannot read anything on the display after first powering up:
1.Observe LED readout.
a.If LED meter reads all eights and periods, go to step 3.
b.If LED meter displays anything else, go to step 2.
2.Press button twice to turn Analyzer OFF and ON again. LED meter should now read all eights and periods. Go to step 3.
Teledyne Analytical Instruments |