Teledyne 3000PA operating instructions Percent Oxygen Analyzer Installation

Page 27

Percent Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

Primary Input Power: The power cord receptacle and fuse block are located in the same assembly. Insert the female plug end of the power cord into the power cord receptacle.

CAUTION: Power is applied to the instrument's circuitry as long as the instrument is connected to the power source. The switch on the front panel is for switching power on or off to the displays and out- puts only.

The universal power supply requires a 85–250 V ac, 47-63 Hz power source.

Fuse Installation: The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. The fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. (See Fuse Replacement in chapter 5, maintenance.)

Analog Outputs: There are four DC output signal connectors with spring terminals on the panel. (Two of them are optional, as explained below.) There are two wires per output with the polarity noted. See Figure 3-

4. The outputs are:

0-1 V dc % Range:

Voltage rises linearly with increasing oxygen con-

 

centration, from 0 V at 0 percent to 1 V at full scale.

 

(Full scale = 100% of programmable range.)

0-1 V dc Range ID:

0.25 V = Low Range, 0.5 V = Medium Range,

0.75 V = High Range, 1 V = Air Cal Range.

4-20 mA dc % Range: –M option only. Current increases linearly with increasing oxygen concentration, from 4 mA at 0 percent to 20 mA at full scale. (Full scale = 100% of programmable range.)

4-20 mA dc Range ID: –M option only. 8 mA = Low Range, 12 mA = Medium Range, 16 mA = High Range, 20 mA = Air Cal Range.

Teledyne Analytical Instruments

3-5

Image 27
Contents Model 3000PA Teledyne Analytical Instruments Model 3000PAIii Percent Oxygen Analyzer Specific Model InformationInstrument Serial Number Includes the following options Operation Model 3000PA Table of Contents IntroductionInstallation OperationalTheoryMaintenance Percent Oxygen AnalyzerAppendix Combustible GAS Usage Warning Percent Oxygen Analyzer Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction Model 3000PA Model DesignationsModel 3000PA Front Panel Front Panel Operator InterfaceIntroduction Model 3000PA Recognizing Difference Between LCD Rear Panel Equipment InterfaceAnalog Outputs Optional RS-232 PortRemote Probe NetworkPrinciples of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Percent Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3000PA+ 2H O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsMicro-Fuel Cell Class Class A-3 CellClass A-5 Cell Class B-3 Cell Hydrogen and/or Helium ServiceSample System Class B-1 CellPiping Layout and Flow Diagram for Standard Model Span Zero Sample Electronics and Signal ProcessingBlock Diagram of the Model 3000P Electronics Operational Theory Model 3000PA Percent Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Rear Panel Connections Installation Model 3000PARear Panel of the Model 3000PA Gas ConnectionsElectrical Connections Percent Oxygen Analyzer Installation Teledyne Analytical Instruments Percent Oxygen Analyzer Installation Command Description Remote Probe Connector Parameter SettingTesting the System Installing the Micro-Fuel CellAuto Ranging on Percent Oxygen Analyzer OperationOperation Model 3000PA Using the Data Entry and Function ButtonsHierarchy of Functions and Subfunctions System FunctionSetting the Display Password Protection Percent Oxygen Analyzer Operation Setting up an Auto-CalEntering the Password Installing or Changing the Password Logout Version Screen Cell Failure Span FunctionsSpan Cal Auto Mode SpanningManual Mode Spanning Alarms FunctionInstallation AL-1 AL-2 Choose Alarm Range Function Setting the Analog Output RangesFixed Range Analysis Operation Model 3000PA Autoranging AnalysisSignal Output Voltage Signal Current Signal Output V dc Output mA dcAnalyze Function Range Voltage Current mA Storing and Handling Replacement Cells Routine MaintenanceCell Replacement Percent Oxygen Analyzer MaintenanceWhen to Replace a Cell Maintenance Model 3000PRemoving the Micro-Fuel Cell IndexRemoving or Installing a Micro-Fuel Cell Cell Warranty System Self Diagnostic Test Fuse ReplacementPreamp PowerMajor Internal Components AnalogRear-Panel Screws CleaningTroubleshooting SolutionPossible Cause Maintenance Model 3000P Teledyne Analytical Instruments Percent Oxygen Analyzer Appendix Model 3000PA SpecificationsAppendix Model 3000PA Qty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListSeries Analyzers Calibration GAS Conversions Zero Cal ZeroSettlingMan ENT To Begin