Percent Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
4.3.5 System
The Model 3000PA has a
The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at will. To initiate a self diagnostic test during operation:
Press the System button to start the System function.
Contrast Function is DISABLED | Contrast |
(Refer to Section 1.6) | PSWD Logout More |
Use the < > arrow keys to blink More, then press Enter.
Version
Cell Output: ### µA
Use the < > arrow keys again to move the blinking to the
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right corner of the screen. When the testing is complete, the results are displayed.
Power: OK Analog: OK
Preamp: 3
The module is functioning properly if it is followed by OK. A number indicates a problem in a specific area of the instrument. Refer to Chapter 5 Maintenance and Troubleshooting for
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
4.3.6 Version Screen
Move the < > arrow key to More and press Enter. With Version blinking, press Enter. The screen displays the manufacturer, model, and software version information.
Teledyne Analytical Instruments |